Fat distribution analysis in salmon fillets using non-contact near infrared
interactance imaging: a sampling and calibration strategy Vegard H. Segtnan, Martin Høy, Frank Lundby, Bjørg Narum and Jens Petter
Wold* Nofima Food, Matforsk AS, Osloveien 1, N-1430 Ås, Norway. E-mail: jens.petter.wold@nofima.no
ABSTRACT:
An online NIR interactance imaging
instrument was tested for fat distribution analysis in raw and salted salmon fillets. Approximately 3000 spectra were collected for each fillet when passing under the instrument on a
conveyor belt (approximately 1 s exposure). The instrument was calibrated using five cylindrical plugs (15 mm diameter) from each fillet. The fat content was measured for each of
these plugs using 1H-NMR spectrometry and the spectra from each plug region were averaged and used for calibration and validation. It was found that online NIR
interactance imaging is well-suited for distributional fat analysis in raw and salted intact salmon fillets. The local sampling and calibration strategy using 15 mm diameter plugs for
reference analysis and spectral averaging was found to provide relevant information and robust models. The average prediction errors (root mean square error of cross-
validation) for raw and salted fillets in combination were approximately 2% fat for local plug regions.
Keywords: remote interactance, online, NIR imaging, salmon, sampling
Back to Table of Contents |