Near infrared applications in the quality control of seed cotton Antonio J. Gaitán-Jurado,a María García-Molina,b Francisco Peña-Rodrígueza and Victor
Ortiz-Somovillaa aIFAPA-Centro 'Alameda del Obispo' (Postharvest and Food Technology Area), Avda. Menéndez Pidal s/n, Apdo.
Correos 3.092, 14080 Cordoba, Spain. E-mail: antonioj.gaitan.ext@juntadeandalucia.es bCITAGRO, Parque Tecnológico de Andalucía, C/
Marie Curie, nš4, local D1, 29590 Malaga, Spain
ABSTRACT:
In the cotton industry the most important parameters for quality of seed cotton (unginned, harvested cotton) are moisture
content and impurity level. In Spain, these quality parameters are regulated and cotton producers are awarded incentives. This study was performed to investigate the potential of
near infrared (NIR) spectroscopy for determining moisture content and impurity level of seed cotton. A total of 100 cotton samples collected from eight cotton ginners were used.
Two spectrometers were employed, one a research level instrument (Foss NIRSystems 6500), the other with a reduced wavelength range more suitable for use at-line (InfraXact
Lab). Moisture content was measured using a moisture meter, which is used in cotton industry, and by a drying oven technique. Impurity level was determined by gravimetric
analysis. Partial least square regression and classification were performed for moisture and impurity level of cotton, respectively. In sample preparation, the NIRSystems 6500 with
a back cover of the sample cell was better in repeatability of spectra than the InfraXact Lab with no back cover. For moisture, the best calibration was developed by using the
reference values from the oven drying method and using the first derivative and standard normal variate and detrending as pre-processing of the spectral data, giving a standard
error of 0.44% in the range of 5.83–14.96% moisture content. The classification model for the impurity level was developed with a combination of first derivative and
multiplicative scatter correction of the spectral data and resulted in an accuracy of 80%. For both moisture and impurity parameters, using the NIRSystems 6500 instrument gave
slightly better results than using the InfraXact Lab.
Keywords: cotton, quality control, near infrared spectroscopy, moisture content, impurity levels