Abstract

European Journal of Mass Spectrometry
Volume 15 Issue 2, Pages 249–260 (2009)
doi: 10.1255/ejms.967

Development of an ion trap/multi-turn time-of-flight mass spectrometer with potential-lift

Kenichi Iwamoto,a Hirofumi Nagaob and Michisato Toyodac
aDepartment of Chemistry, Graduate School of Science, Osaka Prefecture University, 1-1 Gakuencho Nakaku, Sakai, Osaka 599- 8531, Japan
bDivision of Sustainable Energy and Environmental Engineering, Graduate School of Engineering, Osaka University, 2-1 Yamadaoka, Suita, Osaka 565-0781, Japan
cDepartment of Physics, Graduate School of Science, Osaka University, 1-1 Machikaneyama, Toyonaka, Osaka 560-0043, Japan

An ion trap/multi-turn time-of-flight (ToF) mass spectrometer with potential-lift has been developed. This system consists of an external ion source, a lens system, an ion trap, a potential-lift, a multi-turn ToF mass spectrometer and a detector. The ion trap consists of hyperbolic electrode cross-sections (Paul trap) and is used as an ion storage device. The potential-lift, which is part of the flight tube, was attached between the ion trap and the multi-turn ToF mass spectrometer. The potential-lift is known to be useful for increasing the kinetic energy of the ions. In order to check the ability of the potential-lift, mass distributions of [(CsI)n Cs]+ clusters (n = 1–9) were measured. The relative intensity ratios of the [(CsI)nCs]+ clusters were consistent with the results obtained using other apparatus. To check the properties of the new apparatus, Xe+ isotopes were analyzed using either a linear or multi-turn ToF mass spectrometer. In the linear mode, the mass resolution was 500. In the multi-turn mode, the resolution depended on the number of cycles of the multi-turn ToF mass spectrometer; the mass resolution was 4400 (FWHM) after nine cycles. This new apparatus with a high resolution will be useful for measurements of ion–molecule reactions and photodissociations.

Keywords: ion trap, time-of-flight mass spectrometry, multi-turn, potential-lift


Full-text article (1241 kB) (subscribers only)

Buy article on-line for £20 (get immediate access)

Alerting Service

 RSS Feed

Permalink: http://dx.doi.org/10.1255/ejms.967
QR Code (what is this?):


Alerting Services

Our Table of Contents Alerting Service will keep you up-to-date with the latest research published in our journals.

You can also follow our journals on Twitter or subscribe to their RSS feeds.  Follow us on Twitter and Subscribe to our RSS Feeds

Sign Up Now

Subscriptions

Discover the benefits of subscribing to our periodicals

  • Quality Science
  • Fair Pricing
  • Important Research
  • Flexible Subscriptions

Subscribe Today

Sample Copy of EJMS