A comparison between matrix-assisted laser desorption/ionisation time-of-flight
mass spectrometry and size exclusion chromatography in the mass characterisation of synthetic polymers with narrow molecular-mass distributions: Poly(methyl methacrylate) and
poly(styrene) Paul M. Lloyd, Kevin G. Suddaby, Julie E. Varney, Elaine Scrivener, Peter J. Derrick* and David M.
Haddleton* Department of Chemistry and Institute of Mass Spectrometry, University of Warwick, Coventry CV4 7AL, UK
ABSTRACT:
A series of poly(methyl
methacrylate)s and poly(styrene)s (masses 1000–12,000 Da) with narrow molecular-mass distributions have been analysed by matrix-assisted laser desorption/ionisation
time-of flight (MALDI-TOF) mass spectrometry and size exclusion chromatography (SEC).The resultant number average Mn, mass (weight) average
Mw, and z average Mz molecular masses as determined from the two techniques are presented and compared. These data show
MALDI-TOF mass spectrometry and size exclusion chromatography (SEC) to be in good agreement, typically to within 10% for the analyses of these types of polymers.
Keywords:
Back to Table of Contents |